FDA Device recall Z-1578-2026
DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028…
- FDA Device
- Class I
- ongoing
- Published 2026-04-01
On 2026-04-01 the FDA classified a Class I recall of DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028… by Merit Medical Systems, citing 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/throm….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1578-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-02-13 |
| Published | 2026-04-01 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Merit Medical Systems |
| Distribution | US |
Product
DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A, H787103028061/A, H787103028075/A, H787103028085/A, H787103028095/A, H787103028165/A, H787103028181/A, H787103028185/A, H787103028191/A, H787103028195/A, H787103028205/A, H787103028215/A, H787103028225/A, H787103028235/A
Reason
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
Identifiers
| code_info | REF:UDI-DI/Lot: H787103028015/A:00884450794524/I3049562; H787103028025/A:00884450794531I3309371; H787103028031/A:008844…REF:UDI-DI/Lot: H787103028015/A:00884450794524/I3049562; H787103028025/A:00884450794531I3309371; H787103028031/A:00884450794548/I2929299, I3072553, I3154760, I3175731, I3178921, I3315957; H787103028035/A:00884450786130/I2910259, I2910261, I2910262, I2980545, I2995795, I3021388, I3076851, I3085062, I3219446, I3219454, I3255314, I3288649, I3334618; H787103028041/A:00884450794555/I2909647, I3115465, I3178909, I3236154; H787103028045/A:00884450786161/I3004072, I3015761, I3044845, I3072554, I3108153, I3176646, I3178918, I3213147, I3219456, I3281054, I3334627; H787103028051/A:00884450794562/I2995730, I3072556, I3213145, I3219455, I3288560, I3302443; H787103028055/A:00884450794579/I3178903); H787103028061/A:00884450794586/I2929300, I2995814, I3072557, I3165105, I3207168, I3236155, I3334619; H787103028075/A:00884450794609/I2995817), I3209313, I3270826; H787103028085/A:00884450794623/I2890956, I3072561, I3108155, I3288562, I3302444; H787103028095/A:00884450794630/I2909667, I2929315, I3021151, I |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1578-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.