RecallRadar

FDA Device recall Z-1575-2013

Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. Contents: 81.4-cm catheter with length markers and guide wire, 15 T-gauge introduce…

On 2013-07-03 the FDA classified a Class I recall of Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. Contents: 81.4-cm catheter with length markers and guide wire, 15 T-gauge introduce… by Medtronic Neuromodulation, citing the Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1575-2013
ClassificationClass I
Statusterminated
Initiated2013-06-03
Published2013-07-03
Last updated2026-09-13 11:53 UTC
CompanyMedtronic Neuromodulation
DistributionUS

Product

Medtronic INDURA 1P Intrathecal Catheter, model 8709SC. Contents: 81.4-cm catheter with length markers and guide wire, 15 T-gauge introducer needle, Sutureless pump connector with attached 7.6-cm catheter and Connector pin, Transparent Strain-relief sleeves, and anchors. Contents of inner package are STERILE. The implantable Medtronic INDURA 1P Model 8709SC Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8709SC catheter. The catheter connects to the pump at the catheter port.

Reason

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.

Identifiers

code_infoProduct having a Use By Date prior to 25 Aug 2014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1575-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.