FDA Device recall Z-1574-2020
1. Groshong 9.5 Dual-Lumen CV Catheter, Ref: 7726950, UDI: (01)00801741041570
- FDA Device
- Class II
- terminated
- Published 2020-04-01
On 2020-04-01 the FDA classified a Class II recall of 1. Groshong 9.5 Dual-Lumen CV Catheter, Ref: 7726950, UDI: (01)00801741041570 by Bard Peripheral Vascular, citing firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1574-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-10-01 |
| Published | 2020-04-01 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Bard Peripheral Vascular |
| Distribution | US |
Product
1. Groshong 9.5 Dual-Lumen CV Catheter, Ref: 7726950, UDI: (01)00801741041570; 2. Groshong 9.5 F Dual-Lumen CV Catheter, REF: 7726954, UDI: (01)00801741041587
Reason
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Identifiers
| code_info | 7726950 Lot # RECQ1106, RECR2195 7726954 Lot # RECU1837 |
|---|
Status history