FDA Device recall Z-1572-2026
MICS3 Angled Sagittal Saw Attachment
- FDA Device
- Class II
- ongoing
- Published 2026-03-25
On 2026-03-25 the FDA classified a Class II recall of MICS3 Angled Sagittal Saw Attachment by Howmedica Osteonics, citing A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Sa….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1572-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-02-02 |
| Published | 2026-03-25 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Howmedica Osteonics |
| Distribution | US |
Product
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Reason
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Identifiers
| code_info | Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL; |
|---|