FDA Device recall Z-1569-2016
enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Tr…
- FDA Device
- Class II
- terminated
- Published 2016-05-04
On 2016-05-04 the FDA classified a Class II recall of enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Tr… by Ortho Clinical Diagnostics, citing the electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migra….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1569-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-03-17 |
| Published | 2016-05-04 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | AZ, CA, CO, DE, FL, GA, ID, MD, MO, MS, NC, NJ, NY, OH, PA, TN, WV |
Product
enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System -- NOTE: Outside the US the system is also known as VITROS Automation Solution (VAS) TC Automation System (configured with Thermo Scientific Recapper Module from VITROS Automation Solutions), Product Code 6844300 --- The enGen (TM) Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TC Automation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen (TM) System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. The enGen (TM) Laboratory Automation System contains a module (i.e. enGen (TM)Laboratory Automation Recapper Module) that applies caps to the sample collection container after the sample has been aspirated. In this module, an electrical cable supplies power to the cap vibratory feeder and is supposed to be located above the vibratory feeder support plate. The cap vibratory feeder sits upon a feeder support plate. Un-capped sample containers pass underneath this support plate prior to being re-capped for storage. The Recapper Module is supplied by Thermo Fisher Scientific.
Reason
The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.
Identifiers
| code_info | US Serial Numbers: JFBX395J, GBX395J, J953252, JJXBQW4J, J3RHC23J, J86PWC5J, J5LQ6Q2J, JFCM073J, J54FC01J, J5J4B45J, J7…US Serial Numbers: JFBX395J, GBX395J, J953252, JJXBQW4J, J3RHC23J, J86PWC5J, J5LQ6Q2J, JFCM073J, J54FC01J, J5J4B45J, J7KSC15J, J4HX7S4J, JGWXQ52J, J953221, JG7BFB5J, JHPKG25J, JCXL941J, J736BM4J, JFXCBG3J; -- Foreign Serial Numbers: J89133 (VAS), JHWJZY4J, J3J4B45J, J953244, JCT20Q3J, J16XQH4J, GWJVY4J, DWX052J, J271264, J271368, J271371, J271269, J85206, J953216, DP2LF62, J7023874, J7023847, J7025017, J7024893, J7023645, J7023435, J57663, J800103, S/N 89VQ762 (VAS), J75863, J75691, J30372, J3055503, J3055496. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1569-2016%22