RecallRadar

FDA Device recall Z-1566-2023

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 126…

On 2023-05-24 the FDA classified a Class I recall of Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 126… by Icu Medical, citing due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery ru….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1566-2023
ClassificationClass I
Statusongoing
Initiated2023-03-22
Published2023-05-24
Last updated2026-09-13 11:56 UTC
CompanyIcu Medical
Distributionn/a

Product

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973

Reason

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Identifiers

code_infoAll devices containing CSB BATTERY displaying the first two (2) characters as 22 or lower