RecallRadar

FDA Device recall Z-1566-2019

Maquet Getinge-BEQ-TOP 5909 1/4 PACK W QPID Material: 701064992

On 2019-06-05 the FDA classified a Class II recall of Maquet Getinge-BEQ-TOP 5909 1/4 PACK W QPID Material: 701064992 by Maquet Cardiovascular, citing custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1566-2019
ClassificationClass II
Statusterminated
Initiated2019-04-18
Published2019-06-05
Last updated2026-09-13 11:55 UTC
CompanyMaquet Cardiovascular
DistributionUS

Product

Maquet Getinge-BEQ-TOP 5909 1/4 PACK W QPID Material: 701064992

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Identifiers

code_info
Batch Numbers: 3000043679 3000047252 3000050833 3000053134 3000054956 3000056048 3000060481 3000063710 300006…Batch Numbers: 3000043679 3000047252 3000050833 3000053134 3000054956 3000056048 3000060481 3000063710 3000067185 3000071400 3000073429 3000075387 3000075899

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1566-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.