FDA Device recall Z-1559-2015
Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and gener…
- FDA Device
- Class II
- terminated
- Published 2015-05-06
On 2015-05-06 the FDA classified a Class II recall of Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and gener… by Steris, citing should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning. In….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1559-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-02-19 |
| Published | 2015-05-06 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Steris |
| Distribution | US |
Product
Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and general surgical instruments prior to final disinfection and sterilization as required.
Reason
Should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning. In addition, it is possible that the ultrasonic generator may stop working. The machine does not provide an alert notifying the user of this device malfunction.
Identifiers
| code_info | All units distributed between 12/20/2010 and 02/18/2015 |
|---|