RecallRadar

FDA Device recall Z-1549-2024

Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using th…

On 2024-04-24 the FDA classified a Class II recall of Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using th… by Siemens Healthcare Diagnostics, citing potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1549-2024
ClassificationClass II
Statusongoing
Initiated2024-03-13
Published2024-04-24
Last updated2026-09-13 11:57 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211

Reason

Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.

Identifiers

code_infoUDI-DI: 00630414610849 All lots

Status history