RecallRadar

FDA Device recall Z-1548-2015

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro di…

On 2015-05-06 the FDA classified a Class II recall of Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro di… by Dako North America, citing if a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1548-2015
ClassificationClass II
Statusterminated
Initiated2015-04-01
Published2015-05-06
Last updated2026-09-13 11:53 UTC
CompanyDako North America
DistributionUS

Product

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.

Reason

If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.

Identifiers

code_info
N/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1…N/A Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355