FDA Device recall Z-1548-2014
Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on th…
- FDA Device
- Class II
- terminated
- Published 2014-05-14
On 2014-05-14 the FDA classified a Class II recall of Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on th… by Ge Healthcare, citing medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1548-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-04-02 |
| Published | 2014-05-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ge Healthcare |
| Distribution | n/a |
Product
Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
Reason
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Identifiers
| code_info | HDJR50780 HDJR50781 HDJR56911 HDJR57608 HDJR57609 HDJR57610 HDJR57611 HDJR57636 HDJR57637 HDJR57638 HDJR57692…HDJR50780 HDJR50781 HDJR56911 HDJR57608 HDJR57609 HDJR57610 HDJR57611 HDJR57636 HDJR57637 HDJR57638 HDJR57692 HDJR57693 HDJR57694 HDJR57695 HDJR57696 HDJR57697 HDJR57698 HDJR57699 HDJR57700 HDJR57701 HDJR57702 HDJR57703 HDJR57704 HDJR57705 HDJR57706 HDJR57707 HDJR57784 HDJR57785 HDJR57786 HDJR57787 HDJR57788 HDJR57789 HDJR57790 HDJR57791 HDJR57792 HDJR57793 HDJR57794 HDJR57795 HDJR57925 HDJR57926 HDJR57935 HDJR57936 HDJR57937 HDJR57938 HDJR57939 HDJR57944 HDJR57948 HDJR57949 HDJR57950 HDJR57951 HDJR57952 HDJR57953 HDJR57954 HDJR57955 HDJR57956 HDJR57957 HDJR57980 HDJR57981 HDJR57985 HDJR57986 HDJR57987 HDJR57988 HDJR58013 HDJR58014 HDJR58015 HDJR58039 HDJR58040 HDJR58041 HDJR58046 HDJR58047 HDJR58076 HDJR58082 HDJR58086 HDJR58087 HDJR58088 HDJR58091 HDJR58092 HDJR58093 HDJR58094 HDJR58095 HDJR58096 HDJR58097 HDJR58098 HDJR58099 HDJR58100 HDJR58101 HDJR58102 HDJR58103 HDJR58104 HDJR58105 HDJR58106 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1548-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.