RecallRadar

FDA Device recall Z-1546-2014

Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), cons…

On 2014-05-14 the FDA classified a Class I recall of Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), cons… by Alere San Diego, citing test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a sign….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1546-2014
ClassificationClass I
Statusterminated
Initiated2014-04-16
Published2014-05-14
Last updated2026-09-13 11:53 UTC
CompanyAlere San Diego
DistributionUS

Product

Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).

Reason

Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Identifiers

code_info
Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/…Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/28/14, including the following: 324810, 324811, 324812, 327543, 327544, 327549, 328109, 328705, 328707, 330238, 330852, 331340, 331916, 331922, 332841, 332842, 332843, 332959, 333362, 333654, 333655, 333864, 333865, 334572, 334575, 335552, 335553, 335796, 335797, 336426, 336668, 337350, 338965, 338966, 339257, 339707, 327856V Expiration date, or Expected shelf life: range from 4/30/14 to 1/31/15

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1546-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.