FDA Device recall Z-1545-2026
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices
- FDA Device
- Class I
- ongoing
- Published 2026-03-18
On 2026-03-18 the FDA classified a Class I recall of Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices by Trividia Health, citing the system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended ac….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1545-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-02-06 |
| Published | 2026-03-18 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Trividia Health |
| Distribution | US |
Product
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43, TRUE METRIX Kit/RE4H01-51, TRUE METRIX Meter Only (mg/dL)/RE4i01-00, TRUE METRIX Kit (mg/dL)/RE4i01-01, TRUE METRIX Kit (mg/dL)/RE4i01-02, TRUE METRIX Starter Kit (mg/dL)/RE4i03-02, TRUE METRIX Kit (mmol/L)/RE4i03-11, Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00, Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00, TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00, TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11, TRUE METRIX Kit (mg/dL)/RE4i61-02, TRUE METRIX Kit (mg/dL)/RE4i61-02THI, TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11, TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12.
Reason
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Identifiers
| code_info | Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE…Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940. International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/ |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1545-2026%22