GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317
- FDA Device
- Class II
- ongoing
- Published 2025-04-16
On 2025-04-16 the FDA classified a Class II recall of GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317 by Agilent Technologies, citing DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1544-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-03-20 |
| Published | 2025-04-16 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Agilent Technologies |
| Distribution | MA |
Product
GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).
Reason
DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.
Identifiers
| code_info | UDI-DI: 05700571112601. REF/Lot: K1201-64105/0006798023, 5190-7317/0006793917 |
|---|