RecallRadar

FDA Device recall Z-1544-2025

GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317

On 2025-04-16 the FDA classified a Class II recall of GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317 by Agilent Technologies, citing DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1544-2025
ClassificationClass II
Statusongoing
Initiated2025-03-20
Published2025-04-16
Last updated2026-09-13 11:57 UTC
CompanyAgilent Technologies
DistributionMA

Product

GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).

Reason

DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.

Identifiers

code_infoUDI-DI: 05700571112601. REF/Lot: K1201-64105/0006798023, 5190-7317/0006793917

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1544-2025%22