FDA Device recall Z-1544-2021
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
- FDA Device
- Class I
- ongoing
- Published 2021-05-19
On 2021-05-19 the FDA classified a Class I recall of Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part by Infusion Pump Repair, citing bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1544-2021 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2021-03-26 |
| Published | 2021-05-19 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Infusion Pump Repair |
| Distribution | CA |
Product
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Reason
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Identifiers
| code_info | Devices distributed between 01/15/2021 and 03/14/2021. Sale Order #/Serial #: 1599930872/ 12335834, 13617863, 1344726…Devices distributed between 01/15/2021 and 03/14/2021. Sale Order #/Serial #: 1599930872/ 12335834, 13617863, 13447266, 13089105, 12333728, 13616451, 13618575, 12729374; 82943/ 4122252, 4122320; 01112101A/ 12638236, 13813590, 12639090, 13744738, 9951014 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.