FDA Device recall Z-1541-2014
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors)
- FDA Device
- Class II
- terminated
- Published 2014-05-14
On 2014-05-14 the FDA classified a Class II recall of RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors) by Med Mizer, citing all configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holdin….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1541-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-04-18 |
| Published | 2014-05-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Med Mizer |
| Distribution | US |
Product
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Reason
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th
Identifiers
| code_info | Model 762: 300000-300191 Model 763: 200000 Model 763A: 400000-40103 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1541-2014%22