RecallRadar

FDA Device recall Z-1539-2023

LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, SMALL Item Number: 15-2220/40

On 2023-05-10 the FDA classified a Class II recall of LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, SMALL Item Number: 15-2220/40 by Linkbio, citing additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased we….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1539-2023
ClassificationClass II
Statusongoing
Initiated2023-03-21
Published2023-05-10
Last updated2026-09-13 11:56 UTC
CompanyLinkbio
DistributionNV, NY

Product

LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, SMALL Item Number: 15-2220/40

Reason

Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected

Identifiers

code_infoUDI-DI: 04026575429073 All lot numbers within expiration date

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1539-2023%22