RecallRadar

FDA Device recall Z-1536-2014

Anti EpCam PE - Analyte Specific Reagent

On 2014-05-07 the FDA classified a Class II recall of Anti EpCam PE - Analyte Specific Reagent by Bd Biosciences Systems And Reagents, citing two lots of Anti EpCAM PE-Analyte Specific Reagent are mislabeled - the actual antibody present in the vial is CD9 PE.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1536-2014
ClassificationClass II
Statusterminated
Initiated2014-04-04
Published2014-05-07
Last updated2026-09-13 11:53 UTC
CompanyBd Biosciences Systems And Reagents
DistributionAZ, CA, FL, MI, MO, NY, TX

Product

Anti EpCam PE - Analyte Specific Reagent. Anti-EpCAM is directed against human epithelial adhesion molecule that has been identified as a 40 -kilodalton (kd) protein in Western Blotting.

Reason

Two lots of Anti EpCAM PE-Analyte Specific Reagent are mislabeled - the actual antibody present in the vial is CD9 PE.

Identifiers

code_infoCatalog number 347211; Lot number 3295974 - expiry date 2015-09-30; Lot number 3259822 - expiry date 2015-09-30.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.