RecallRadar

FDA Device recall Z-1536-2013

90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is int…

On 2013-06-19 the FDA classified a Class II recall of 90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is int… by Orthopediatrics, citing post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1536-2013
ClassificationClass II
Statusterminated
Initiated2013-04-19
Published2013-06-19
Last updated2026-09-13 11:53 UTC
CompanyOrthopediatrics
DistributionUS

Product

90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.

Reason

Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.

Identifiers

code_infoAll lot numbers are affected. Catalog No. 00-1200-1002

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1536-2013%22