FDA Device recall Z-1523-2018

Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester…

On 2018-05-02 the FDA classified a Class II recall of Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester… by Sofradim Production, citing there have been patient reports of abdominal hernia recurrence following hernia repair. The recalling firm has added a statement to the device's Instructions for Use to make surge….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1523-2018
ClassificationClass II
Statusterminated
Initiated2018-02-23
Published2018-05-02
Last updated2026-09-13 11:54 UTC
CompanySofradim Production
DistributionUS

Product

Versatex Monofilament Mesh 50 x 50cm, Product No. VTX5050M The device is made out of a macroporous three-dimensional monofilament polyester textile. Its intended use is for the repair of abdominal wall hernias or other fascial deficiencies that require the addition of a reinforcing material.

Reason

There have been patient reports of abdominal hernia recurrence following hernia repair. The recalling firm has added a statement to the device's Instructions for Use to make surgeons aware of risks in using the device in Transversus Abdominis Muscle Release (TAR) procedures.

Identifiers

code_infoUDI - 10884521532298 Lots SPC1091X SPJ0706X SPJ1016X SQA0990X SQI0302X SQJ0443X SQK0319X SRA2188X SRI0548X

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1523-2018%22