FDA Device recall Z-1522-2014

Integra CUSA Excel Disposable Wrench

On 2014-05-07 the FDA classified a Class II recall of Integra CUSA Excel Disposable Wrench by Integra Lifesciences, citing some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1522-2014
ClassificationClass II
Statusterminated
Initiated2014-03-10
Published2014-05-07
Last updated2026-09-13 11:53 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessory used with the CUSA Excel+ System ultrasonic surgical instrument. The torque wrench is used to properly secure the CUSA tip to the CUSA handpiece in the sterile field.

Reason

Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.

Identifiers

code_infoProduct/Catalog Number C5601; Lot Numbers 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873 and 1135014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.