FDA Device recall Z-1520-2013

DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty

On 2013-06-19 the FDA classified a Class II recall of DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty by Depuy Orthopaedics, citing the Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1520-2013
ClassificationClass II
Statusterminated
Initiated2013-05-03
Published2013-06-19
Last updated2026-09-13 11:53 UTC
CompanyDepuy Orthopaedics
DistributionUS

Product

DePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used in reverse shoulder arthroplasty. The orientation guide is used to properly align the glenosphere.

Reason

The Glenosphere Orientation Guide Instrument Used with the Delta XTEND Reverse Shoulder is being recalled because an arrow that is etched on the instrument was incorrectly placed on the opposite side of the instrument. Using the improperly etched arrow as a guide could potentially lead to incorrect orientation of the glenosphere implant.

Identifiers

code_infoPart Number 230795000/Lot Number 5120443

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.