FDA Device recall Z-1519-2014
VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc
- FDA Device
- Class II
- terminated
- Published 2014-05-07
On 2014-05-07 the FDA classified a Class II recall of VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc by Ortho Clinical Diagnostics, citing ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Special….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1519-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-03-26 |
| Published | 2014-05-07 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY
Reason
Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.
Identifiers
| code_info | Lot F3168 (exp. 30 Sep 2014) |
|---|