FDA Device recall Z-1518-2014
IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diag…
- FDA Device
- Class II
- terminated
- Published 2014-05-07
On 2014-05-07 the FDA classified a Class II recall of IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diag… by Siemens Healthcare Diagnostics, citing siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1518-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-12-20 |
| Published | 2014-05-07 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | n/a |
Product
IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.
Reason
Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.
Identifiers
| code_info | Lots 260 through 275 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1518-2014%22