RecallRadar

FDA Device recall Z-1518-2014

IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diag…

On 2014-05-07 the FDA classified a Class II recall of IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diag… by Siemens Healthcare Diagnostics, citing siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1518-2014
ClassificationClass II
Statusterminated
Initiated2013-12-20
Published2014-05-07
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
Distributionn/a

Product

IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.

Reason

Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.

Identifiers

code_infoLots 260 through 275

Status history