ExacTrac 6.0 is a patient positioning and monitoring system

On 2014-05-07 the FDA classified a Class II recall of ExacTrac 6.0 is a patient positioning and monitoring system by Brainlab Ag, citing when using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, ther….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1513-2014
ClassificationClass II
Statusterminated
Initiated2014-02-24
Published2014-05-07
Last updated2026-09-13 11:53 UTC
CompanyBrainlab Ag
DistributionCA, IN, LA, MA, OH, TX, WI

Product

ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED

Reason

When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.

Identifiers

code_infoExacTrac version v.6.0.3.