FDA Device recall Z-1512-2014

GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and pa…

On 2014-05-07 the FDA classified a Class I recall of GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and pa… by Hospira, citing there are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1512-2014
ClassificationClass I
Statusterminated
Initiated2014-02-18
Published2014-05-07
Last updated2026-09-13 11:53 UTC
CompanyHospira
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

GemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and patient bolus cord. The docking station is also used for hosting the GemStar pump for general infusions. List Numbers: 13075-XX-XX.

Reason

There are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List 13000-XX, 13100-XX, or 13150-XX) the potential exists for the GemStar Phase 3 pump to fail to power up while connected to the Docking Station, and 2) when either a GemStar Phase 3 (List 13000-XX, 13100-XX, or 13150-XX)or GemStar Phase 4 pump (List 13086-XX, 13087-XX, or 13088-XX) is used in conjunction with both a Docking Station and an External Battery Pack accessory (List 13073-XX) there exists a possibility that the GemStar pump will display error code 11/003 and give an audible alarm indicating excessive input voltage from the external sources. If the GemStar pump detects what is perceived to be more than 3.6 Volts measured on the external voltage input, the pump will stop the infusion and alarm both with an audible sound as well as a visual alarm.

Identifiers

code_info
*** US DOCKING STATIONS *** List Numbers: 13075-04-01, 13075-04-03, 13075-04-05, 13075-04-07, 13075-04-09. ***…*** US DOCKING STATIONS *** List Numbers: 13075-04-01, 13075-04-03, 13075-04-05, 13075-04-07, 13075-04-09. *** FOREIGN DOCKING STATIONS *** List Numbers: 13075-24-01, 13075-24-03, 13075-24-05, 13075-24-07, 13075-92-09.