On 2022-08-17 the FDA classified a Class II recall of The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglyce… by Alere San Diego, citing cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations gr….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1508-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-06-24 |
| Published | 2022-08-17 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Alere San Diego |
| Distribution | AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MN, MO, NC, NE, NV, NY, OH, OK, PA, RI, SC, TN, TX, VA |
Product
The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987
Reason
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Identifiers
| code_info | All lots of Part/UDI-DI: 10-991/00893038002654, 99021/00899722002344, 14-531/04571226470831, 10-990/00893038002647,…All lots of Part/UDI-DI: 10-991/00893038002654, 99021/00899722002344, 14-531/04571226470831, 10-990/00893038002647, 10-989/00893038002630, 10-987/00893038002616 |
|---|