FDA Device recall Z-1505-2017
Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator
- FDA Device
- Class II
- terminated
- Published 2017-03-29
On 2017-03-29 the FDA classified a Class II recall of Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator by Ohmeda Medical, citing certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power co….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1505-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-11-18 |
| Published | 2017-03-29 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ohmeda Medical |
| Distribution | US |
Product
Giraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment for the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Reason
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Identifiers
| code_info | See Attached Spreadsheat |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1505-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.