RecallRadar

FDA Device recall Z-1504-2015

TS Herbst Nuts with Backing (10/pk) and TS Herbst Nuts Universal (10/pk) A Herbst Appliance is used to affect changes in the skeletal relat…

On 2015-04-29 the FDA classified a Class II recall of TS Herbst Nuts with Backing (10/pk) and TS Herbst Nuts Universal (10/pk) A Herbst Appliance is used to affect changes in the skeletal relat… by Specialty Appliance Works, citing the Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1504-2015
ClassificationClass II
Statusterminated
Initiated2015-03-09
Published2015-04-29
Last updated2026-09-13 11:53 UTC
CompanySpecialty Appliance Works
DistributionUS

Product

TS Herbst Nuts with Backing (10/pk) and TS Herbst Nuts Universal (10/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.

Reason

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.