RecallRadar

FDA Device recall Z-1502-2017

Giraffe Incubator The Giraffe Incubator is an infant incubator

On 2017-03-29 the FDA classified a Class II recall of Giraffe Incubator The Giraffe Incubator is an infant incubator by Ohmeda Medical, citing certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1502-2017
ClassificationClass II
Statusterminated
Initiated2016-11-18
Published2017-03-29
Last updated2026-09-13 11:54 UTC
CompanyOhmeda Medical
DistributionUS

Product

Giraffe Incubator The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed temperature controlled environment

Reason

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Identifiers

code_infoSee Attached Spreadsheat

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1502-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.