FDA Device recall Z-1501-2016

CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuou…

On 2016-04-27 the FDA classified a Class II recall of CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuou… by Medtronic Minimed, citing medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time referenc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1501-2016
ClassificationClass II
Statusterminated
Initiated2016-02-19
Published2016-04-27
Last updated2026-09-13 11:54 UTC
CompanyMedtronic Minimed
DistributionMN

Product

CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuous glucose information

Reason

Medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their sensor data were subject to the Central Standard Time zone. The CareLink iPro2 Professional CGM (continuous glucose monitor) data uploading error that resulted in an inaccurate time zone being applied to 19 CareLink iPro reports.

Identifiers

code_infon/a