FDA Device recall Z-1492-2020

UroDiagnost Eleva, system code 708033

On 2020-03-25 the FDA classified a Class II recall of UroDiagnost Eleva, system code 708033 by Philips North America, citing A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1492-2020
ClassificationClass II
Statusterminated
Initiated2019-12-30
Published2020-03-25
Last updated2026-09-13 11:55 UTC
CompanyPhilips North America
DistributionUS

Product

UroDiagnost Eleva, system code 708033

Reason

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Identifiers

code_infoSerial numbers 64 48 2 60

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.