RecallRadar

FDA Device recall Z-1491-2026

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The…

On 2026-03-11 the FDA classified a Class II recall of Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The… by Diagnostica Stago, citing the potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1491-2026
ClassificationClass II
Statusongoing
Initiated2026-02-04
Published2026-03-11
Last updated2026-09-13 11:58 UTC
CompanyDiagnostica Stago
DistributionUS

Product

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable

Reason

The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

Identifiers

code_info
Model/Catalog Number (REF): 00516; UDI: (1) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (2) (01)0360745…Model/Catalog Number (REF): 00516; UDI: (1) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (2) (01)03607450005165(11)250221(17)260531(10)272859(241)00516, (3) (01)03607450005165(11)250306(17)260630(10)273018(241)00516, (4) (01)03607450005165(11)250430(17)260831(10)273426(241)00516; Lot numbers: (1) 271 971, (2) 272 859, (3) 273 018, (4) 273 426; Lot Code: 271 971 (01)03607450005165(11)241001(17)260131(10)271971(241)00516

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1491-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.