FDA Device recall Z-1487-2014

iLab Ultrasound Imaging System

On 2014-04-30 the FDA classified a Class II recall of iLab Ultrasound Imaging System by Boston Scientific, citing specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview dista….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1487-2014
ClassificationClass II
Statusterminated
Initiated2014-04-03
Published2014-04-30
Last updated2026-09-13 11:53 UTC
CompanyBoston Scientific
DistributionUS

Product

iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.

Reason

Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.

Identifiers

code_info
Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 751…Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.