RecallRadar

FDA Device recall Z-1484-2025

GE Healthcare Centricity PACS-IW (PACS-IW), Model Numbers: 1) 2042988-001

On 2025-04-09 the FDA classified a Class II recall of GE Healthcare Centricity PACS-IW (PACS-IW), Model Numbers: 1) 2042988-001 by Ge Healthcare, citing there is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1484-2025
ClassificationClass II
Statusongoing
Initiated2025-02-19
Published2025-04-09
Last updated2026-09-13 11:57 UTC
CompanyGe Healthcare
Distributionn/a

Product

GE Healthcare Centricity PACS-IW (PACS-IW), Model Numbers: 1) 2042988-001; 2) 2049587-009; 3) 2049587-011; 4) 2049587-012; 5) 2049587-015; 6) 2049588-012 ; 7) 2052829-001.

Reason

There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.

Identifiers

code_info
Model/Catalog/Code # 2042988-001, Installed Product ID 707509-Centricity PACS-IW-01452391, UDI UDI Not Required, System…Model/Catalog/Code # 2042988-001, Installed Product ID 707509-Centricity PACS-IW-01452391, UDI UDI Not Required, System ID HCIT4250187IW; Model/Catalog/Code # 2042988-001, Installed Product ID 613493-Centricity PACS-IW-02292544, UDI UDI Not Required, System ID HCIT4507688PACS; Model/Catalog/Code # 2049587-009, Installed Product ID 4276-1-Centricity PACS-IW-00558274, UDI UDI Not Required, System ID PACS-IW-168; Model/Catalog/Code # 2049587-011, Installed Product ID 12830-1-Centricity PACS-IW-00558312, UDI UDI Not Required, System ID PACS-IW-477; Model/Catalog/Code # 2049587-011, Installed Product ID 4352-1-Centricity PACS-IW-00558316, UDI UDI Not Required, System ID PACS-IW-89; Model/Catalog/Code # 2049587-012, Installed Product ID 100858-1-Centricity PACS-IW-00558426, UDI UDI Not Required, System ID PACS-IW-412; Model/Catalog/Code # 2049587-012, Installed Product ID 4294-1-Centricity PACS-IW-00558379, UDI UDI Not Required, System ID PACS-IW-304; Model/Catalog/Code # 2049587-012,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1484-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.