FDA Device recall Z-1484-2013
Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and…
- FDA Device
- Class II
- terminated
- Published 2013-06-12
On 2013-06-12 the FDA classified a Class II recall of Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and… by Zimmer Surgical, citing this removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1484-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-05-07 |
| Published | 2013-06-12 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer Surgical |
| Distribution | US |
Product
Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities
Reason
This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.
Identifiers
| code_info | Serial Number 500191 , 500192,500193, 500194,500196,500197,500198,500199, 500200, 500201,500202,500203, 500204, 500205,…Serial Number 500191 , 500192,500193, 500194,500196,500197,500198,500199, 500200, 500201,500202,500203, 500204, 500205,500206, 500207,500208,500209,500210, 500211,500212, 500213,500215,500216,500217, 500219,500220, 500221,500222, 500223, 500224, 500225,500226,500227,500228,500229,500230, 500231,500232, 500233,500234,500235,500236, 500237, 500238,500239,500240,500241 , 500242, 500243,500244,500245,500246,500247,500248,500249, 500250, 500251,500252,500253,500254, 500255,500256,500257,500258,500259, 500260,500261,500262,500263,500264, 500265,500266,500267,500268,500269, 500270,500291, 500292,500293,500295,500296,500297,500298, 500299,500300,500301 , 500302,500303, 500304, 500306, 500307,500308,500309,500310, 500331 , 500332, 500334, 500335, 500336, 500337, 500338, 500339, 500340, 500341,500342, 500343, 500344, 500345, 500346, 500347, 500348,500349, 500350, 500351 , 500352, 500353, 500354,500355, 500357, 500358, 500359, 500368, 500369,500370,500371 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1484-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.