RecallRadar

FDA Device recall Z-1483-2013

The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and wid…

On 2013-06-12 the FDA classified a Class II recall of The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and wid… by Zimmer Surgical, citing this removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1483-2013
ClassificationClass II
Statusterminated
Initiated2013-05-07
Published2013-06-12
Last updated2026-09-13 11:53 UTC
CompanyZimmer Surgical
DistributionUS

Product

The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities

Reason

This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Identifiers

code_info
Serial Number 100211, 100338, 100515, 100560, 100647, 100755, 101104, 101345, 101562, 101769, 101890, 101983, 102560, 1…Serial Number 100211, 100338, 100515, 100560, 100647, 100755, 101104, 101345, 101562, 101769, 101890, 101983, 102560, 103407, 103509, 103816, 104529, 104863, 105049, 105072, 105604, 106394, 106474, 106624, 107861, 108103, 108203, 108210, 108339, 108911, 109445, 109543, 110034, 110837, 111046, 111047, 111048, 111049, 111050, 111051, 111052, 111053, 111054, 111055, 111056, 111057, 111058, 111059, 111060, 111061, 111062, 111063, 111064, 111065, 111066, 111067, 111068, 111069, 111071, 111072, 111073, 111074, 111075, 111076, 111079, 111081, 111083, 111084, 111085, 111086, 111087, 111089, 111090, 111091, 111092, 111096, 111100, 111101, 111102, 111104, 111105, 111116, 111121, 111122

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1483-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.