FDA Device recall Z-1479-2026
Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product…
- FDA Device
- Class II
- ongoing
- Published 2026-03-11
On 2026-03-11 the FDA classified a Class II recall of Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product… by Olympus, citing ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1479-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-14 |
| Published | 2026-03-11 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Olympus |
| Distribution | US |
Product
Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Reason
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
Identifiers
| code_info | Model/Catalog Number: SPL-G; UDI: 00821925044203; Serial Numbers: All; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1479-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.