On 2014-04-30 the FDA classified a Class II recall of LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel… by Hospira, citing hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a bur….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1479-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-02-25 |
| Published | 2014-04-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hospira |
| Distribution | FL, IA, IN, KS, KY, ME, MO, MT, PA, TN, TX, UT, VA, WA |
Product
LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.
Reason
Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.
Identifiers
| code_info | Product List Number: 112410403; Lot Numbers: 280055H, 341005H |
|---|