FDA Device recall Z-1477-2021
Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model…
- FDA Device
- Class II
- terminated
- Published 2021-04-28
On 2021-04-28 the FDA classified a Class II recall of Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model… by Smiths Medical Asd, citing specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. T….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1477-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-03-31 |
| Published | 2021-04-28 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Smiths Medical Asd |
| Distribution | US |
Product
Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.
Reason
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
Identifiers
| code_info | a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 40…a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1477-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.