FDA Device recall Z-1477-2017

C2 CryoBalloon Ablation System - Controller Cap

On 2017-03-22 the FDA classified a Class II recall of C2 CryoBalloon Ablation System - Controller Cap by C2 Therapeutics, citing the Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controll….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1477-2017
ClassificationClass II
Statusterminated
Initiated2017-01-06
Published2017-03-22
Last updated2026-09-13 11:54 UTC
CompanyC2 Therapeutics
DistributionUS

Product

C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

Reason

The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.

Identifiers

code_info
All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from…All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from February 1 2016 through February 8, 2017. All numbers below have been updated. Lot numbers: Expiry dates 01122016-02 12-Jul-16; 06242016-01 1-Jan-2017; 11282016-03 2-Jun-2017; 01042016-03 14-Jul-16; 06292016-02 6-Jan-2017; 09202016-01 21-Mar-2017; 07062016-01 8-Jan-2017; 09222016-01 23-Mar-2017; 01252016-02 02-Aug-16; 07082016-02 13-Jan 2017; 09232016-01 28-Mar-2017; 01282016-01 01-Aug-16; 07082016-03 14-Jan-2017; 09232016-02 27-Mar-2017; 01292016-01 02-Aug-16; 07152016-01 21-Jan-2017; 09292016-01 4-Apr-2017; 02022016-03 09-Aug-16; 07152016-02 20-Jan-2017; 09302016-03 4-Apr-2017; 02082016-01 09-Aug-16; 07212016-01 25-Jan-2017; 10052016-02 19-Apr-2017; 02092016-02 11-Aug-16; 07252016-01 2

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1477-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.