SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100

On 2023-05-10 the FDA classified a Class II recall of SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100 by Stryker, citing fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1474-2023
ClassificationClass II
Statusongoing
Initiated2023-03-02
Published2023-05-10
Last updated2026-09-13 11:56 UTC
CompanyStryker
DistributionUS

Product

SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200

Reason

Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.

Identifiers

code_info
Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159…Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 1689