FDA Device recall Z-1474-2020

CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress

On 2020-03-25 the FDA classified a Class I recall of CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress by Cme America, citing infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1474-2020
ClassificationClass I
Statusterminated
Initiated2020-01-07
Published2020-03-25
Last updated2026-09-13 11:55 UTC
CompanyCme America
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY

Product

CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress

Reason

Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than expected drug delivery when infusing at high flow rates (greater-than 500 mL/h)

Identifiers

code_infoAll serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1474-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.