FDA Device recall Z-1472-2023

Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatri…

On 2023-05-03 the FDA classified a Class II recall of Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatri… by Draeger Medical Systems, citing the software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this conditio….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1472-2023
ClassificationClass II
Statusongoing
Initiated2023-04-05
Published2023-05-03
Last updated2026-09-13 11:56 UTC
CompanyDraeger Medical Systems
DistributionUS

Product

Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,

Reason

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Identifiers

code_infoUDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower