FDA Device recall Z-1472-2023
Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatri…
- FDA Device
- Class II
- ongoing
- Published 2023-05-03
On 2023-05-03 the FDA classified a Class II recall of Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatri… by Draeger Medical Systems, citing the software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this conditio….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1472-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-04-05 |
| Published | 2023-05-03 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Draeger Medical Systems |
| Distribution | US |
Product
Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,
Reason
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Identifiers
| code_info | UDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1472-2023%22