FDA Device recall Z-1471-2023

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical…

On 2023-05-03 the FDA classified a Class II recall of Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical… by Draeger Medical Systems, citing the software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this conditio….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1471-2023
ClassificationClass II
Statusongoing
Initiated2023-04-05
Published2023-05-03
Last updated2026-09-13 11:56 UTC
CompanyDraeger Medical Systems
DistributionUS

Product

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Reason

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Identifiers

code_infoUDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1471-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.