RecallRadar

FDA Device recall Z-1468-2026

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

On 2026-03-04 the FDA classified a Class II recall of Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167 by Medline Industries, citing affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1468-2026
ClassificationClass II
Statusongoing
Initiated2026-01-08
Published2026-03-04
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Reason

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Identifiers

code_infoUDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1468-2026%22