RecallRadar

FDA Device recall Z-1466-2026

Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

On 2026-03-04 the FDA classified a Class II recall of Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A by Medline Industries, citing affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1466-2026
ClassificationClass II
Statusongoing
Initiated2026-01-08
Published2026-03-04
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

Reason

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Identifiers

code_info
SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GB…SKU DYNDA2705 UDI-DI (ea) 10193489468632 UDI-DI (case) 40193489468633 Kit lots 24DBL787 24HBR630 25ABH699 25DBV209 25GBQ405 SKU DYNJ63347A UDI-DI (ea) 10193489278064 UDI-DI (case) 40193489278065 Kit lots 24AMC874 24AME139 24AMH820 24BMF097 24CMF691 24GMB769 24GMK036 24IMC124 24LMA110 25AMJ634 25GMD386 25IMB510 25JMI708

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1466-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.