On 2026-03-04 the FDA classified a Class II recall of Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200 by Medline Industries, citing affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1463-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-01-08 |
| Published | 2026-03-04 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200
Reason
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Identifiers
| code_info | IR PACK UDI-DI (ea) 10195327621780 UDI-DI (case) 40195327621781 Kit lots 24BBP627 24DBO029 24FBI148 24JBA858 24KBD925 2…IR PACK UDI-DI (ea) 10195327621780 UDI-DI (case) 40195327621781 Kit lots 24BBP627 24DBO029 24FBI148 24JBA858 24KBD925 24LBG978 25ABQ002 25EBV162; NO CATHETER DIALYSIS INSERTION UDI-DI (ea) 10195327675219 UDI-DI (case) 40195327675210 Kit Lots 24EBM915 24GBC287 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1463-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.