RecallRadar

FDA Device recall Z-1463-2018

The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Descr…

On 2018-05-02 the FDA classified a Class II recall of The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Descr… by Integra Lifesciences, citing incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1463-2018
ClassificationClass II
Statusterminated
Initiated2017-12-20
Published2018-05-02
Last updated2026-09-13 11:54 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm

Reason

Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required.

Identifiers

code_info
TRL092002506 Lot Numbers: 19419-1 15916-1 15457-1 201274-1 201893-1 2046177 TRL092002507 Lot Numbers: 1941…TRL092002506 Lot Numbers: 19419-1 15916-1 15457-1 201274-1 201893-1 2046177 TRL092002507 Lot Numbers: 19419-2 15916-2 15457-2 201274-2 201893-2 2046178 TRL092002508 Lot Numbers: 19419-3 15916-3 15457-3 201274-3 201893-3 2046179 TRL092002509 Lot Numbers: 19419-4 15916-4 15457-4 201274-4 201893-4 20461710 TRL092002510 Lot Numbers: 19419-5 15916-5 15457-5 201274-5 201893-5 20461711 TRL092002511 Lot Numbers: 19419-6 15916-6 15457-6 201274-6 201893-6 20461712 TRL092002512 Lot Numbers: 19419-7 15916-7 15457-7 201274-7 201893-7 20461713 TRL092002513 Lot Numbers: 19419-8 15916-8 15457-8 201274-8 201893-8 20461714 TRL092002514 Lot Numbers: 19419-9 15916-9 15457-9 201274-9 201893-9 20461715 TRL092002515 Lot Numbers: 19419-10 15916-10 15457-10 201274-10 201893-10 20461716 TRL092002516 Lot Numbers: 19419-11 15916-11 15457-11 201274-11 201893-11 20461717

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1463-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.