RecallRadar

FDA Device recall Z-1461-2015

TIGERPAW System II, Part number C-TP-1507 (7 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ost…

On 2015-04-29 the FDA classified a Class I recall of TIGERPAW System II, Part number C-TP-1507 (7 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ost… by Laax, citing increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1461-2015
ClassificationClass I
Statusterminated
Initiated2015-03-25
Published2015-04-29
Last updated2026-09-13 11:53 UTC
CompanyLaax
DistributionUS

Product

TIGERPAW System II, Part number C-TP-1507 (7 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).

Reason

Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure of the TigerPaw System II fastener.

Identifiers

code_infoPart number C-TP-1507 (7 connector) All serial numbers/lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1461-2015%22